Stability Test Chamber(UVA)



Humidity: 20~95%RH
Internal Volume: 150L–500L
Power Supply: AC 220V±10%, 50Hz
Temperature Control: Range: 15–65°C
Fluctuation: ±0.5°C
Deviation: ±1.0°C (without light)

Product introduction

Labonce-CGS-FC Series Integrated Drug Stability Test Chamber (Combined Temperature + Humidity + Visible Light + Near-UV Control) Featuring an all-new structural design and high-quality imported components, this system ensures long-term stable operation with reliable performance. 


Features of Stability Test Chamber

1. Humidity Control: Range: 20–95% RH  Fluctuation: ±2.0% RH Deviation: ±3.0% RH (without light)

2. Visible Light Range: Adjustable: 100–8000 LUX Standard: 4500±500 LUX Total illuminance: ≥1.2×10⁶ Lux·hr
3. Near-UV Range: Adjustable: 0.84–5 W/m² Total energy: ≥200 W·hr/m²
4. UV Wavelength: 320–400 nm
5. Control System: Touchscreen controller (centralized control for temperature, humidity, visible light, near-UV).
6. Data Printing: Prints real-time values for all four parameters (temperature, humidity, visible light, near-UV).
7. Temperature & Humidity Controller: 7-inch+ color touchscreen, programmable, with 3-level access control and audit trail.
8. Humidity Sensor: Imported capacitive sensor, direct detection, maintenance-free.
9. Refrigeration Compressor: Inverter-driven fully sealed compressor (imported), long lifespan, low noise, 50%+ energy savings.
10.  Data System: Stores data for ≥5 years; prints model number, temperature, humidity, and illuminance values.
11. Alarm System: On-site alarms (buzzer for deviations in temp/humidity/light). Independent over temperature alarm. Remote SMS alerts (including power failure alerts).
12. Chamber Material: Inner wall: Mirror-finish stainless steel (304). Outer shell: High-quality steel plate with powder coating.
13. Dual Protection: Dual dry-burn prevention system. Independent over-temperature protection (auto power  cutoff).
14. Safety Features: Compressor overheat/over pressure/overload protection. Fan overheat protection.



Specification of Stability Test Chamber

Model
Labonce-150CGS
Labonce-250CGS
Labonce-400CGS
Labonce-500CGS
Temperature Range( ℃ )
15~65
Humidity Range (RH)
20~95%
Capacity (L)

150

250

400

500

Interior Dimensions

W×D×H(mm)

600×405×620

600x500x830

600x700x950

680×680×1080

External Dimensions

W×D×H(mm)

780×830×1480

780x880x1650

780x1060x1750

860×1050×1850

Power (kw)

1.2

1.5

2.0

2.2

Shelf (Standard)

3

3

3

4

Remark

1) It has four functions: temperature, humidity, visible light and near ultraviolet. The illuminance measurement is equipped with visible light and near ultraviolet sensors, which realizes automatic illuminance control, and automatic printing and storage of visible light and near ultraviolet values.

2)Visible light range: 100 ~ 8000Lux, illuminance deviation: 4500 ± 500Lux, the total illuminance of the illumination test shall not be less than 1.2×106 Lux·hr; near ultraviolet range: 0.84 ~ 1.0 w/m2 , the total illuminance of the near ultraviolet shall not be less than 200w·hr/m2.



SEND A MESSAGE

Related products
  • Stability Test Chamber Stability Chamber(>800L)

    Humidity:15~95%RH Capacity:800~1000L Power: AC220V±10% 50Hz/60Hz Environment Temperature:+5  ~  35℃ Temperature Fluctuation<±0.5℃ Temperature Uniformity <±1.0℃ Humidity Fluctuation < ±2%RH Humidity Uniformity <  ±3%RH

  • Comprehensive Stability Chamber Stability Test Chamber (Triple Independent)

    Chamber capacity: 430L-930L Humidity control: Range 20-95%RH Temperature control: Range 15-65°C Power: AC 220V±10%, 50Hz Visible light: 100-6,000 Lux Near-UV: 0.84-1 W/m² UV wavelength: 320-400 nm

  • Temperature Humidity Stability Test Chamber Stability Chamber(3000L)

    Capacity: 3000L Humidity: 15~95%RH Environment Temperature: +5  ~  35℃ Temperature Fluctuation<±0.5℃ Temperature Uniformity <±1.0℃ Humidity Fluctuation < ±2%RH Humidity Uniformity <  ±3%RH Power: AC220V±10% 50Hz/60Hz

  • Stability chamber Stability Test Chamber(Dual Independent)

    Humidity:20~95%RH Internal Volume: 300L-600L Temperature Control Range: 15-65°C Temperature Fluctuation: ±0.5°C Temperature Deviation: ±1.0°C (without light) Humidity Control Range: 20-95%RH Humidity Fluctuation: ±2.0%RH Humidity Deviation: ±3.0%RH Power Supply: AC 220V±10%, 50Hz

  • Stability Chamber, Stability test chamber Stability Chamber(<500L)

    Humidity: 15~95%RH Capacity: 150~500L Temperature Fluctuation: <±0.5℃ Temperature Uniformity: <±1.0℃ Humidity Fluctuation: < ±2%RH Humidity Uniformity: <  ±3%RH Power: AC220V±10% 50Hz/60Hz Environment Temperature: +5  ~  35℃

  • Stability Test Chamber(Special design) Stability Test Chamber(Special design)

    Humidity:10~90%RH Chamber capacity: Customized Temperature control: Customized Humidity control: Customized Capacity:Design Power supply: AC 220V±10%, 50Hz Operating environment: +5-35°C

  • High and low temperature alternating chamber Programmable Temperature & Humidity Chamber

    TEMP:-20~150℃;-40~150℃;-60~150℃

  • High Temperature Constant Test Chamber High Temperature Constant Test Chamber

    TEMP:10~100℃ Capacity:250L~2000L

Beijing Labonce Themostatic Technology Company

Professional Technology , Good Products

Email: Sophia@labonce.com FDA/ICH stability Chamber

© 2008-2026 Beijing Labonce Themostatic Technology Company All Rights Reserved

Online contact

Online contact

If you have questions or suggestions,please leave us a message,we will reply you as soon as we can!