stability chamberslanbeishilabonce    
  stability chamber
               
     
stability chamber
       
 
Photo-stability testing chambers
 

1、The latest airway system design achieves uniform of temperature and humidity from different parts inside the chamber;Interior material is mirror stainless steel SUS304, no pollution source, easy for clean.

2、Industrial compressor is European original imported completely sealed for high efficiency , low noise, assure machine of long time continuous running .

3、Temperature and humidity integrated sensor is European original imported for high precise, low floating annually. It can test humidity directly , no need of maintenance.

4、Original imported temperature and humidity controller for fast reaction, low deviation of system, high precise control.

5、Standard meets : ICH-2003 Q1(A) CP2010

 

Name

Stability Chambers(GMP)

Stability Chambers (GMP)

Model

Labonce-120CGS

Labonce-250CGS

Labonce-350CGS

Labonce-700CGS

Temperature Range

0~65℃(Illumination:15~65℃)

10~65℃(Illumination:15~65℃)

Accuracy

±0.5℃(Illumination:±1.0℃)

±0.5℃(Illumination:±1.0℃)

Deviation

±1.5℃(Illumination:±2.0℃)

±1.5℃(Illumination:±2.0℃)

Humidity Range

40~95%RH

40~95%RH

Accuracy

±3%RH

±3%RH

Deviation

±4%RH

±4%RH

Illumination intension

0~10000LUX

0~10000LUX

Deviation

±500LUX

±500LUX

Control manner

Temperature and humidity balance Control

Compressor

Original Imported industry compressor ( European Union)

Humidity Sensor

Original Imported humidity sensor ( European Union)

Controller

Original Imported Digital Display Controller(JAPAN)

Data Recorder

Optional

Test point

40℃ 75%RH、25℃ 60%RH

Temperature alarm

High and low boundary deviation alarm for temperature(Optional)

Humidity alarm

High and low boundary deviation alarm for humidity(Optional)

Dual protections

Independent overheat protection warning system

Operating temp.

+5~35℃

Internal material

Mirror surface Stainless steel SUS304

External material

Steel powder coated

Water tank

External Water Tank

Power source

AC 220V±10% 50HZ

Safety device

Compressor and blower overheat protection system, Over temperature protection, overpressure and overload protection for compressor, Water scarcity protection

Max Power(kW)

1.5

2.5

2.5

3.0

Capacity (L)

150

250

350

700

Interior dimensions
(mm)W×D×H

600×405×620

680×500×730

680×530×1100

820×800×1070

External dimensions
(mm)W×D×H

760×830×1480

840×870×1550

840×900×1750

1100×1280×1820

Trays (standard)

3

3

4

4

Optional

1、Data record:U disk(with microprinter) or Recorder with paper(with chart drive)
2、Dual stage refrigeration system
3、Alarm: Temperature & Humidity deviation alarm
43Q validation documents
5Calibration certificate

Remark

Please confirm whether the external dimension of this model chamber can be smooth for elevator、corridor or gate, whether there is enough space for it in lab.